Skip to content
CVE PROGRAM SOURCE RECORD10.0 Critical

CVE-2024-48967

Life2000 ventilator and Service PC lack sufficient audit logging capabilities

Description

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

Affected products

Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.

Baxter Life2000 Ventilation System

Default version status: unaffected

  • 06.08.00.00 and prioraffected

References