{"id":"CVE-2024-48967","state":"PUBLISHED","title":"Life2000 ventilator and Service PC lack sufficient audit logging capabilities","description":"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.","publishedAt":"2024-11-14T21:40:44.904Z","updatedAt":"2024-11-15T14:28:13.605Z","severity":{"score":10,"label":"Critical","version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","source":"Baxter"},"assigner":"Baxter","products":["Baxter Life2000 Ventilation System"],"affected":[{"vendor":"Baxter","product":"Life2000 Ventilation System","defaultStatus":"unaffected","versions":[{"version":"06.08.00.00 and prior","status":"affected"}]}],"references":["https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"],"weaknesses":["CWE-778"],"ingestedAt":"2026-10-29T20:07:56.516Z","source":{"repository":"CVEProject/cvelistV5","path":"cves/2024/48xxx/CVE-2024-48967.json","commit":"556c9d3a8cd17e218bb8e452ba7cd619a384ff47"},"raw":{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2024-48967","state":"PUBLISHED","datePublished":"2024-11-14T21:40:44.904Z","dateUpdated":"2024-11-15T14:28:13.605Z","assignerShortName":"Baxter","assignerOrgId":"dba971b9-eb30-4121-91e1-3b45611354aa","dateReserved":"2024-10-10T19:24:34.436Z"},"containers":{"cna":{"title":"Life2000 ventilator and Service PC lack sufficient audit logging capabilities","descriptions":[{"lang":"en","value":"The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance."}],"affected":[{"vendor":"Baxter","product":"Life2000 Ventilation System","defaultStatus":"unaffected","versions":[{"version":"06.08.00.00 and prior","status":"affected"}]}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":10,"baseSeverity":"CRITICAL","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-778","description":"CWE-778: Insufficient Logging","lang":"en","type":"CWE"}]}],"providerMetadata":{"shortName":"Baxter","orgId":"dba971b9-eb30-4121-91e1-3b45611354aa","dateUpdated":"2024-11-14T21:52:23.702Z"},"datePublic":"2024-11-14T21:38:00.000Z","impacts":[{"capecId":"CAPEC-117","descriptions":[{"lang":"en","value":"CAPEC-117 Interception"}]},{"capecId":"CAPEC-441","descriptions":[{"lang":"en","value":"CAPEC-441 Malicious Logic Insertion"}]}],"source":{"discovery":"UNKNOWN"},"x_generator":{"engine":"Vulnogram 0.2.0"}},"adp":[{"title":"CISA ADP Vulnrichment","affected":[{"vendor":"baxter","product":"life2000_ventilator_firmware","defaultStatus":"unknown","versions":[{"version":"0","status":"affected","lessThan":"6.08.00.00","versionType":"custom"}],"cpes":["cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"]}],"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2024-11-15T14:28:10.676566Z","id":"CVE-2024-48967","options":[{"Exploitation":"none"},{"Automatable":"yes"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"providerMetadata":{"shortName":"CISA-ADP","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","dateUpdated":"2024-11-15T14:28:13.605Z"}}]}}}