Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

Description

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected products

Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.

Medtronic MiniMed 508 pump

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 511 pump

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 512/712 pumps

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 712E pump

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 515/715 pumps

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 522/722 pumps

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 522K/722K pumps

Default version status: unaffected

  • All versionsaffected

Medtronic MiniMed Paradigm 523/723 pumps

Default version status: unaffected

  • 0 to ≤ Software Versions 2.4Aaffected

Medtronic MiniMed Paradigm 523K/723K pumps

Default version status: unaffected

  • 0 to ≤ Software Versions 2.4Aaffected

Medtronic MiniMed Paradigm Veo 554/754 pumps

Default version status: unaffected

  • 0 to ≤ Software Versions 2.6Aaffected

Medtronic MiniMed Paradigm Veo 554CM/754CM pumps

Default version status: unaffected

  • 0 to ≤ Software Versions 2.7Aaffected