Description
In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
Affected products
Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
Default version status: unaffected
1.7.1.47affected01.07.01.53unaffected
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App
Default version status: unaffected
4.5.15.4affected4.5.18unaffected