Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Affected products
Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.
Contec Health CMS8000 Patient Monitor
Default version status: unaffected
All versionsaffected
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01 ↗
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ↗
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor ↗