Skip to content
CVE PROGRAM SOURCE RECORD9.3 Critical

CVE-2024-48974

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

Affected products

Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.

Baxter Life2000 Ventilation System

Default version status: unaffected

  • 06.08.00.00 and prioraffected

References