Skip to content
CVE PROGRAM SOURCE RECORD9.3 Critical

CVE-2019-6538

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Affected products

Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.

Medtronic Conexus Radio Frequency Telemetry Protocol

Default version status: unaffected

  • All versionsaffected

Medtronic MyCareLink Monitor

Default version status: unaffected

  • 24950affected
  • 24952affected

Medtronic CareLink Monitor

Default version status: unaffected

  • 2490Caffected

Medtronic CareLink 2090 Programmer

Default version status: unaffected

  • All versionsaffected

Medtronic Amplia CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Claria CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Compia CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Concerto CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Concerto II CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Consulta CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Evera ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Maximo II CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Maximo II ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Mirro ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Nayamed ND ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Primo ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Protecta ICD, Protecta CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Secura ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Virtuoso ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Virtuoso II ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Visia AF ICD

Default version status: unaffected

  • All versionsaffected

Medtronic Viva CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Brava CRT-D

Default version status: unaffected

  • All versionsaffected

Medtronic Mirro MRI ICD

Default version status: unaffected

  • All versionsaffected

References