Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Affected products
Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.
Dräger Infinity Acute Care System
Default version status: unknown
0 to ≤ VG4.1.1affected0 to ≤ VG4.0.3affectedVG4.2unaffected
Dräger Standalone Infinity M540 patient monitor
Default version status: unknown
0 to ≤ VG4.1.1affectedVG4.2unaffected