Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Affected products
Version statuses are supplied by the assigning organization. Consult the original record for detailed ranges and platform conditions.
Hospira LifeCare PCA Infusion System
Default version status: unaffected
0 to ≤ 5.0affected7.0unaffected
References
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01 ↗
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm ↗
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/ ↗
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json ↗
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 ↗