{"id":"CVE-2025-41690","state":"PUBLISHED","title":"Endress+Hauser: Proline 10 Maintenance credentials may be exposed under certain conditions","description":"A low-privileged attacker in bluetooth range may be able to access the password of a higher-privilege user (Maintenance) by viewing the device’s event log. This vulnerability could allow the Operator to authenticate as the Maintenance user, thereby gaining unauthorized access to sensitive configuration settings and the ability to modify device parameters.","publishedAt":"2025-09-02T08:12:13.946Z","updatedAt":"2025-09-02T13:48:59.186Z","severity":{"score":7.4,"label":"High","version":"3.1","vector":"CVSS:3.1/AV:A/AC:L/PR:L/UI:R/S:U/C:H/I:H/A:H","source":"CERTVDE"},"assigner":"CERTVDE","products":["Endress+Hauser Promag 10 with HART","Endress+Hauser Promag 10 with IO-Link","Endress+Hauser Promag 10 with Modbus","Endress+Hauser Promass 10 with HART","Endress+Hauser Promass 10 with IO-Link","Endress+Hauser Promass 10 with Modbus"],"affected":[{"vendor":"Endress+Hauser","product":"Promag 10 with HART","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promag 10 with IO-Link","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.02","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promag 10 with Modbus","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with HART","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with IO-Link","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.02","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with Modbus","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]}],"references":["https://certvde.com/en/advisories/VDE-2025-068"],"weaknesses":["CWE-532"],"ingestedAt":"2026-10-29T20:09:09.358Z","source":{"repository":"CVEProject/cvelistV5","path":"cves/2025/41xxx/CVE-2025-41690.json","commit":"556c9d3a8cd17e218bb8e452ba7cd619a384ff47"},"raw":{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2025-41690","state":"PUBLISHED","datePublished":"2025-09-02T08:12:13.946Z","dateUpdated":"2025-09-02T13:48:59.186Z","assignerShortName":"CERTVDE","assignerOrgId":"270ccfa6-a436-4e77-922e-914ec3a9685c","dateReserved":"2025-04-16T11:17:48.309Z"},"containers":{"cna":{"title":"Endress+Hauser: Proline 10 Maintenance credentials may be exposed under certain conditions","descriptions":[{"lang":"en","value":"A low-privileged attacker in bluetooth range may be able to access the password of a higher-privilege user (Maintenance) by viewing the device’s event log. This vulnerability could allow the Operator to authenticate as the Maintenance user, thereby gaining unauthorized access to sensitive configuration settings and the ability to modify device parameters."}],"affected":[{"vendor":"Endress+Hauser","product":"Promag 10 with HART","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promag 10 with IO-Link","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.02","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promag 10 with Modbus","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with HART","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with IO-Link","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.02","versionType":"semver"}]},{"vendor":"Endress+Hauser","product":"Promass 10 with Modbus","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThan":"01.00.06","versionType":"semver"}]}],"references":[{"url":"https://certvde.com/en/advisories/VDE-2025-068"}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":7.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"LOW","scope":"UNCHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:A/AC:L/PR:L/UI:R/S:U/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-532","description":"CWE-532 Insertion of Sensitive Information into Log File","lang":"en","type":"CWE"}]}],"providerMetadata":{"shortName":"CERTVDE","orgId":"270ccfa6-a436-4e77-922e-914ec3a9685c","dateUpdated":"2025-09-02T08:12:13.946Z"},"source":{"advisory":"VDE-2025-068","defect":["CERT@VDE#641830"],"discovery":"UNKNOWN"},"x_generator":{"engine":"Vulnogram 0.1.0-dev"}},"adp":[{"title":"CISA ADP Vulnrichment","metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2025-09-02T13:48:53.162675Z","id":"CVE-2025-41690","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"providerMetadata":{"shortName":"CISA-ADP","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","dateUpdated":"2025-09-02T13:48:59.186Z"}}]}}}