{"id":"CVE-2023-1966","state":"PUBLISHED","title":"CVE-2023-1966","description":"Instruments with Illumina Universal Copy Service v1.x and\nv2.x contain an unnecessary privileges vulnerability. An unauthenticated\nmalicious actor could upload and execute code remotely at the operating system\nlevel, which could allow an attacker to change settings, configurations,\nsoftware, or access sensitive data on the affected product.\n\n\n\n\n\n","publishedAt":"2023-04-28T18:06:51.431Z","updatedAt":"2025-01-16T21:35:30.057Z","severity":{"score":7.4,"label":"High","version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H","source":"icscert"},"assigner":"icscert","products":["Illumina  iScan Control Software","Illumina  MiniSeq Control Software","Illumina  iSeq 100","Illumina  MiSeq Control Software","Illumina  MiSeqDx Operating Software","Illumina  NextSeq 500/550 Control Software","Illumina  NextSeq 550Dx Control Software","Illumina  NextSeq 550Dx Operating Software","Illumina  NextSeq 1000/2000 Control Software","Illumina  NovaSeq 6000 Control Software","Illumina  NovaSeq Control Software"],"affected":[{"vendor":"Illumina ","product":"iScan Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0.0","status":"affected"},{"version":"4.0.5","status":"affected"}]},{"vendor":"Illumina ","product":"MiniSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"2.0 ","status":"affected"}]},{"vendor":"Illumina ","product":"iSeq 100","defaultStatus":"unaffected","versions":[{"version":"All versions","status":"affected"}]},{"vendor":"Illumina ","product":"MiSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0 (RUO Mode)","status":"affected"}]},{"vendor":"Illumina ","product":"MiSeqDx Operating Software","defaultStatus":"unaffected","versions":[{"version":"4.0.1 ","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 500/550 Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 550Dx Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0 (RUO Mode)","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 550Dx Operating Software","defaultStatus":"unaffected","versions":[{"version":"1.0.0 ","status":"affected","lessThanOrEqual":"1.3.1","versionType":"custom"},{"version":"1.3.3 ","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 1000/2000 Control Software","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"1.4.1","versionType":"custom"}]},{"vendor":"Illumina ","product":"NovaSeq 6000 Control Software","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"1.7 ","versionType":"custom"}]},{"vendor":"Illumina ","product":"NovaSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"1.8","status":"affected"}]}],"references":["https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01","https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html"],"weaknesses":["CWE-250"],"ingestedAt":"2026-10-29T20:05:36.141Z","source":{"repository":"CVEProject/cvelistV5","path":"cves/2023/1xxx/CVE-2023-1966.json","commit":"556c9d3a8cd17e218bb8e452ba7cd619a384ff47"},"raw":{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2023-1966","state":"PUBLISHED","datePublished":"2023-04-28T18:06:51.431Z","dateUpdated":"2025-01-16T21:35:30.057Z","assignerShortName":"icscert","assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","dateReserved":"2023-04-10T14:50:41.262Z"},"containers":{"cna":{"title":"CVE-2023-1966","descriptions":[{"lang":"en","value":"Instruments with Illumina Universal Copy Service v1.x and\nv2.x contain an unnecessary privileges vulnerability. An unauthenticated\nmalicious actor could upload and execute code remotely at the operating system\nlevel, which could allow an attacker to change settings, configurations,\nsoftware, or access sensitive data on the affected product.\n\n\n\n\n\n"}],"affected":[{"vendor":"Illumina ","product":"iScan Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0.0","status":"affected"},{"version":"4.0.5","status":"affected"}]},{"vendor":"Illumina ","product":"MiniSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"2.0 ","status":"affected"}]},{"vendor":"Illumina ","product":"iSeq 100","defaultStatus":"unaffected","versions":[{"version":"All versions","status":"affected"}]},{"vendor":"Illumina ","product":"MiSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0 (RUO Mode)","status":"affected"}]},{"vendor":"Illumina ","product":"MiSeqDx Operating Software","defaultStatus":"unaffected","versions":[{"version":"4.0.1 ","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 500/550 Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 550Dx Control Software","defaultStatus":"unaffected","versions":[{"version":"4.0 (RUO Mode)","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 550Dx Operating Software","defaultStatus":"unaffected","versions":[{"version":"1.0.0 ","status":"affected","lessThanOrEqual":"1.3.1","versionType":"custom"},{"version":"1.3.3 ","status":"affected"}]},{"vendor":"Illumina ","product":"NextSeq 1000/2000 Control Software","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"1.4.1","versionType":"custom"}]},{"vendor":"Illumina ","product":"NovaSeq 6000 Control Software","defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"1.7 ","versionType":"custom"}]},{"vendor":"Illumina ","product":"NovaSeq Control Software","defaultStatus":"unaffected","versions":[{"version":"1.8","status":"affected"}]}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01"},{"url":"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html","tags":["vendor-advisory"]}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"PHYSICAL","availabilityImpact":"HIGH","baseScore":7.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"LOW","scope":"CHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-250","description":"CWE-250 Execution with Unnecessary Privileges","lang":"en","type":"CWE"}]}],"providerMetadata":{"shortName":"icscert","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","dateUpdated":"2023-04-28T18:06:51.431Z"},"credits":[{"lang":"en","user":"00000000-0000-4000-9000-000000000000","value":"Illumina reported these vulnerabilities to CISA. "}],"source":{"discovery":"INTERNAL"},"workarounds":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"\n\n<span style=\"background-color: rgb(255, 255, 255);\">Illumina recommends using the </span><a target=\"_blank\" rel=\"nofollow\" href=\"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html\">UCS Vulnerability Instructions Guide</a><span style=\"background-color: rgb(255, 255, 255);\">&nbsp;based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software. </span>\n\n<br>"}],"value":"\nIllumina recommends using the  UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software. \n\n\n"}],"x_generator":{"engine":"Vulnogram 0.1.0-dev"}},"adp":[{"title":"CVE Program Container","references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01","tags":["x_transferred"]},{"url":"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html","tags":["vendor-advisory","x_transferred"]}],"providerMetadata":{"shortName":"CVE","orgId":"af854a3a-2127-422b-91ae-364da2661108","dateUpdated":"2024-08-02T06:05:27.096Z"}},{"title":"CISA ADP Vulnrichment","metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2025-01-16T20:30:42.910219Z","id":"CVE-2023-1966","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"providerMetadata":{"shortName":"CISA-ADP","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","dateUpdated":"2025-01-16T21:35:30.057Z"}}]}}}