{"id":"CVE-2022-4992","state":"PUBLISHED","title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering","description":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.","publishedAt":"2026-06-02T21:06:40.281Z","updatedAt":"2026-06-05T11:26:40.821Z","severity":{"score":8.8,"label":"High","version":"4.0","vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","source":"VulnCheck"},"assigner":"VulnCheck","products":["Dräger Infinity Acute Care System","Dräger Standalone Infinity M540 patient monitor"],"affected":[{"vendor":"Dräger","product":"Infinity Acute Care System","defaultStatus":"unknown","versions":[{"version":"0","status":"affected","lessThan":"VG4.2","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.1.1","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.0.3","versionType":"custom"}]},{"vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","defaultStatus":"unknown","versions":[{"version":"0","status":"affected","lessThan":"VG4.2","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.1.1","versionType":"custom"}]}],"references":["https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf","https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering"],"weaknesses":["CWE-345"],"ingestedAt":"2026-10-29T20:05:26.763Z","source":{"repository":"CVEProject/cvelistV5","path":"cves/2022/4xxx/CVE-2022-4992.json","commit":"556c9d3a8cd17e218bb8e452ba7cd619a384ff47"},"raw":{"dataType":"CVE_RECORD","dataVersion":"5.2","cveMetadata":{"cveId":"CVE-2022-4992","state":"PUBLISHED","datePublished":"2026-06-02T21:06:40.281Z","dateUpdated":"2026-06-05T11:26:40.821Z","assignerShortName":"VulnCheck","assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","dateReserved":"2026-06-02T21:02:24.899Z"},"containers":{"cna":{"title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering","descriptions":[{"lang":"en","value":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality."}],"affected":[{"vendor":"Dräger","product":"Infinity Acute Care System","defaultStatus":"unknown","versions":[{"version":"0","status":"affected","lessThan":"VG4.2","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.1.1","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.0.3","versionType":"custom"}]},{"vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","defaultStatus":"unknown","versions":[{"version":"0","status":"affected","lessThan":"VG4.2","versionType":"custom"},{"version":"0","status":"affected","lessThan":"VG4.1.1","versionType":"custom"}]}],"references":[{"url":"https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering","tags":["third-party-advisory"]}],"metrics":[{"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}],"cvssV4_0":{"attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","subConfidentialityImpact":"NONE","vulnIntegrityImpact":"LOW","subIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED","version":"4.0","baseSeverity":"HIGH","baseScore":8.8,"vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N"}},{"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}],"cvssV3_1":{"version":"3.1","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH","baseSeverity":"HIGH","baseScore":8.6,"vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H"}}],"problemTypes":[{"descriptions":[{"cweId":"CWE-345","lang":"en","description":"CWE-345 Insufficient Verification of Data Authenticity","type":"CWE"}]}],"providerMetadata":{"shortName":"VulnCheck","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","dateUpdated":"2026-06-05T11:26:40.821Z"},"datePublic":"2022-01-24T00:00:00.000Z","source":{"discovery":"INTERNAL"},"x_generator":{"engine":"Vulnogram 1.0.2"}},"adp":[{"title":"CISA ADP Vulnrichment","metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2026-06-03T12:45:44.296578Z","id":"CVE-2022-4992","options":[{"Exploitation":"none"},{"Automatable":"yes"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"providerMetadata":{"shortName":"CISA-ADP","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","dateUpdated":"2026-06-03T12:45:51.502Z"}}]}}}