{"id":"CVE-2021-38396","state":"PUBLISHED","title":"Missing Support Integrity Check for Boston Scientific Zoom Latitude","description":"The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.","publishedAt":"2021-10-04T17:35:21.072Z","updatedAt":"2024-09-16T20:46:57.503Z","severity":{"score":6.5,"label":"Medium","version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L","source":"icscert"},"assigner":"icscert","products":["Boston Scientific ZOOM LATITUDE"],"affected":[{"vendor":"Boston Scientific","product":"ZOOM LATITUDE","versions":[{"version":"Model 3120","status":"affected"}]}],"references":["https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"],"weaknesses":["CWE-353"],"ingestedAt":"2026-10-29T19:48:49.871Z","source":{"repository":"CVEProject/cvelistV5","path":"cves/2021/38xxx/CVE-2021-38396.json","commit":"556c9d3a8cd17e218bb8e452ba7cd619a384ff47"},"raw":{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2021-38396","state":"PUBLISHED","datePublished":"2021-10-04T17:35:21.072Z","dateUpdated":"2024-09-16T20:46:57.503Z","assignerShortName":"icscert","assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","dateReserved":"2021-08-10T00:00:00.000Z"},"containers":{"cna":{"title":"Missing Support Integrity Check for Boston Scientific Zoom Latitude","descriptions":[{"lang":"en","value":"The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."}],"affected":[{"vendor":"Boston Scientific","product":"ZOOM LATITUDE","versions":[{"version":"Model 3120","status":"affected"}]}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01","tags":["x_refsource_MISC"]}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"PHYSICAL","availabilityImpact":"LOW","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"LOW","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L","version":"3.1"}}],"problemTypes":[{"descriptions":[{"cweId":"CWE-353","description":"CWE-353 Missing Support for Integrity Check","lang":"en","type":"CWE"}]}],"providerMetadata":{"shortName":"icscert","dateUpdated":"2021-10-04T17:35:21.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6"},"credits":[{"lang":"en","value":"Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH Münster  University of Applied Sciences, Christian Dresen - FH Münster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."}],"datePublic":"2021-09-30T00:00:00.000Z","source":{"advisory":"ICSMA-21-273-01","defect":["CWE-353"],"discovery":"EXTERNAL"},"workarounds":[{"lang":"en","value":"Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."}],"x_generator":{"engine":"Vulnogram 0.0.9"},"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","DATE_PUBLIC":"2021-09-30T21:02:00.000Z","ID":"CVE-2021-38396","STATE":"PUBLIC","TITLE":"Missing Support Integrity Check for Boston Scientific Zoom Latitude"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"ZOOM LATITUDE","version":{"version_data":[{"version_value":"Model 3120"}]}}]},"vendor_name":"Boston Scientific"}]}},"credit":[{"lang":"eng","value":"Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH Münster  University of Applied Sciences, Christian Dresen - FH Münster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."}],"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."}]},"generator":{"engine":"Vulnogram 0.0.9"},"impact":{"cvss":{"attackComplexity":"LOW","attackVector":"PHYSICAL","availabilityImpact":"LOW","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"LOW","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L","version":"3.1"}},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"CWE-353 Missing Support for Integrity Check"}]}]},"references":{"reference_data":[{"name":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01","refsource":"MISC","url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"}]},"source":{"advisory":"ICSMA-21-273-01","defect":["CWE-353"],"discovery":"EXTERNAL"},"work_around":[{"lang":"en","value":"Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."}]}},"adp":[{"title":"CVE Program Container","references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01","tags":["x_refsource_MISC","x_transferred"]}],"providerMetadata":{"shortName":"CVE","orgId":"af854a3a-2127-422b-91ae-364da2661108","dateUpdated":"2024-08-04T01:37:16.545Z"}}]}}}